Brief introduction
Servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100% ; the infant bed can be pulled out from the side door.

Standard Configure
Hood, infant bed, temperature controller, water tank, air filter, I.V. Pole, stand.

Technical Specification
  1. Power supply : AC220V 50HZ
  2. Power input: 600VA
  3. Air Temperature control range 25 ~ 37
  4. Changing of temperature: d0.5
  5. Uniformity of temperature: d0.8
  6. Temperature rising time: d45min
  7. Noise level: d55dBA
Alarming : Audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater.
  • Over-temp alarm: when the temperature control d38
  • Deviation alarm: ± 3
  • Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
  • Fan alarm: when the fan is blocked, speed d1000rpm or stop working;
  • Power failure alarm: when there is no power or the power cord is disconnected.
  • System alarm: when there is E2ROM malfunction inside of temperature controller.
Package
  • Gross Weight 72Kg
  • Measurement 1085mm× 640mm× 1065m


Jika Anda berminat dengan Produk ini, jangan ragu untuk hubungi kami karena kami selalu siap melayani anda. 
Mobile :   081296374738
Jika Anda menyukai Artikel di blog ini, Silahkan klik disini untuk berlangganan gratis via email, dengan begitu Anda akan mendapat kiriman artikel setiap ada artikel yang terbit di Surya Medilab

0 komentar:

Posting Komentar

 
Support : Creating Website | Johny Template | Mas Template
Copyright © 2011. SURYA MEDILAB - All Rights Reserved
Template Modify by Creating Website
Proudly powered by Blogger